一线运维实战手记 · 云计算 · 数据库 · 故障排查
2026年9月4日

CVE-2019-25723 [MEDIUM] – Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handlin

未分类
📑 文章目录

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.